Squarewave Characteristics; Squarewave Checks - Tabor Electronics 5064 User Manual

50 / 100 / 200 ms/s four channel arbitrary waveform generator
Table of Contents

Advertisement

2074 Offset
Setting
+4.000 V
+1.500 V
0.000 V
-1.500 V
-4.000 V
Squarewave
Characteristics

Squarewave Checks

Amplitude:
Output:
Offset:
Test Procedure
1. Perform Offset Accuracy tests on all channels using Table 6-5
Table 6-5, Offset Accuracy
Error Limits
CH1
4.000 V ±45 mV
1.500 V ±20 mV
0 V ±20 mV
-1.500 V ±20 mV
-4.000 V ±45 mV
This tests the characteristics of the square waveform. It includes
transition times, aberrations and skew between channels. Each
channel has its own set of amplifiers and attenuators and therefore,
the characteristics are tested on each channel separately.
Equipment: Oscilloscope, 50 Ω, 20dB feedthrough attenuator
Preparation:
1. Configure the Oscilloscope follows:
Termination:
Setup:
2. Connect 2074 Channel outputs to the oscilloscope input
Configure the 2074 as follows:
Frequency:
Waveform:
Amplitude:
Output:
Test Procedure
1. Perform Squarewave Characteristics tests on all channels using
Table 6-6
Performance Checks
20 mV
On
As specified in Table 6-5
DMM Reading
CH2
CH3
CH4
50 Ω, 20 dB feedthrough attenuator at the
oscilloscope input
As required for the test
1 MHz
Squarewave
5 V
On
6
Test Procedures
Pass
Fail
6-7