2074
User Manual
Amplitude
Accuracy
Amplitude Accuracy
2074 Amplitude
Setting
10.00 V
5.000 V
500 mV
50.00 mV
Offset Accuracy
Offset Accuracy
6-6
Amplitude accuracy checks tests the accuracy of the output amplifier
and attenuators. Each channel has its own set of amplifiers and
attenuators and therefore, the accuracy is tested on each channel
separately.
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Termination:
Function:
2. Connect 2074 Channel outputs to the DMM input
3. Configure the 2074 as follows:
Frequency:
Output:
Amplitude:
Test Procedure
1.
Perform amplitude Accuracy tests on all channels using Table
6-4
Table 6-4, Amplitude Accuracy, DAC output
Error Limits
3.534 V, ±60 mV
1.767 V, ±42 mV
176.7 mV, ± 5.3 mV
17.67 mV, ± 1.9 mV
Offset accuracy checks tests the accuracy of the offset generators.
Each channel has its own set of offset generators and therefore, the
accuracy is tested on each channel separately.
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Termination:
Function:
2. Connect 2074 Channel outputs to the DMM input
3. Configure the 2074 as follows:
Frequency:
50 Ω feedthrough at the DMM input
ACV
1 kHz
On
As specified in Table 6-4
DMM Reading
CH1
CH2
CH3
50 Ω feedthrough at the DMM input
DCV
1 MHz
CH4
Pass
Fail