Installation

Installation setups

Environmental
requirements
Installation setups
Releasing the
transport lock
22
Thermo Scientific Wellwash Versa User Manual
When you set up your Wellwash Versa, avoid sites with excess dust,
vibrations, strong magnetic fields, direct sunlight, draft, excessive
moisture or large temperature fluctuations. Make sure that:
The working area is flat, dry, clean and vibration-proof, and leave
additional room for cables, covers, and so on
The ambient air is clean and free of corrosive vapors, smoke and
dust
The ambient temperature range is between +10°C (50°F) and +40°C
(104°F)
The humidity is low so that condensation does not occur (relative
humidity is between 10% and 80%, non-condensing).
Caution Do not operate the instrument in an environment where
potentially damaging liquids or gases are present.
This section describes the procedures that must be carried out before
instrument operation.
Caution Leave the instrument to stand for at least three hours before
installing and switching it on, so there is no possibility of condensation
causing a short circuit.
Warning All parts of the instrument that come into contact with
potentially infectious materials must be treated as potentially infectious
areas.
It is advisable to adhere to applicable safety precautions, including the
wearing of powder-free gloves, safety glasses and protective clothing, to
avoid potential infectious disease contamination when performing
cleaning procedures and also when making adjustments to the
instrument.
The instrument comes with one transport lock.
Caution Make sure that the transport lock has been released before you
put the instrument into operation.
To release the transport lock:
1. Remove the padded packing material protecting the wash head arm
and plate carrier to expose the transport lock and transport lock tag
(Figure 3–7).