Chapter 2: Installation

Warning!
For more information on main parts of the instrument, refer to Chapter 3: "Wellwash Main Parts".

Unpacking

How to unpack

Move the packed instrument to its site of operation. To prevent condensation, the instrument should
be left in its protective plastic wrapping until the ambient temperature has been reached. Unpack the
Wellwash instrument and accessories carefully with the arrows on the transport package pointing
upwards. Place the instrument onto a laboratory bench.
Caution!
the instructions. Doing so may cause misalignment and will void the instrument warranty.
Retain the original packaging for future transportation. The packaging is designed to assure safe
transport and minimize transit damage. Using other packaging materials may invalidate the warranty.
Also retain all instrument-related documentation provided by the manufacturer for future use.
If you relocate your instrument or ship it for service, refer to "How to pack for service".

Checking delivery for completeness or damage

Check the enclosed packing list against the order. Visually inspect the transport packaging, the
instrument and the accessories for any possible transport damage. If any parts are missing or
damaged, contact your local Thermo Fisher Scientific representative or Thermo Fisher Scientific Oy.
Caution!

Environmental requirements

When you set up your Wellwash, avoid sites with excess dust, vibrations, strong magnetic fields,
direct sunlight, draft, excessive moisture or large temperature fluctuations. Make sure that:
The working area is flat, dry, clean and vibration-proof, and leave additional room for cables,
covers, and so on.
There is at least 10 cm of free space around the instrument on the laboratory bench for
ventilation.
There is sufficient room behind the instrument to enable disconnecting the device.
The ambient air is clean and free of corrosive vapors, smoke and dust.
The ambient temperature range is between +10°C (50°F) and +40°C (104°F).
The humidity is low so that condensing does not occur (relative humidity is between 10% and
80%, non-condensing).
Caution!
or gases are present.
The Wellwash weighs 8 kg [17.6 lbs.] and care must be taken when lifting it.
Do not touch or loosen any screws or parts other than those specifically designated in
If the instrument has been mechanically damaged, ship it for service.
Do not operate the instrument in an environment where potentially damaging liquids
10