Set Up and Run a Scan Method

Exercise 1. Set up a full-scan acquisition method

Exercise 1. Set up a full-scan acquisition method
This exercise changes the default method and saves it as a new method. This
exercise consists of the following tasks:

Task 1. Enter LC acquisition parameters

Steps
1 Display the Method and Run Control
view.
2 Open the method DEF_LC.M.
3 Save the method under a new name,
SULFA MS SCAN 1.M.
1
L
4 Enter a volume of
µ
injection.
18
"Task 1. Enter LC acquisition parameters"
"Task 2. Enter MS acquisition parameters"
Detailed Instructions
• In the view selection area in the
lower left of the ChemStation
window, click Method and Run
Control.
a Select File > Load > Method.
b If necessary, navigate to
C:\CHEM32\1\METHODS.
c Select DEF_LC.M and click OK.
a Select File > Save As > Method.
b In the dialog box, for Name, type
SULFA MS SCAN 1.M.
c Click OK.
d In the box for Comment for method
history, type a comment.
e Click OK.
for the
a Click Set up Instrument Method on
the Instrument menu to open the
Setup Method dialog box.
b Click the HiP Sampler tab.
c In the Injection volume box, type 1
for a 1-µL injection.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
on page 18
on page 20
Comments
• You save the method now with a
new name to avoid inadvertently
overwriting the default method
later.