Steps
2 Enter parameters for the spray
chamber of the ion source:
•
Drying gas flow: 9 L/min
•
Nebulizer pressure: 40 psig
•
Drying gas temperature: 300
•
Capillary voltage: 3000 V
For 6150 with Agilent Jet Stream
technology:
•
Drying gas flow: 7 L/min
•
Nebulizer pressure: 30 psi
•
Drying gas temperature: 350°C
•
Sheath gas temperature: 360 C°
•
Sheath gas flow: 12 L/min
•
Capillary voltage: 4000 V
•
Nozzle voltage: 2000 V
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Detailed Instructions
a Right-click the MSD
icon on the system
diagram and select
Spray Chamber.
o
C
b Enter the parameters described in
step 2
and shown in the figure
below.
c Click OK.
For all models except 6150 with Agilent Jet Stream technology
For 6150 with Agilent Jet Stream technology
Set Up and Run a Scan Method
Task 2. Enter MS acquisition parameters
Comments
2
21