Task 5. Set up the calibration curve

Task 5. Set up the calibration curve

In this exercise, you integrate the rest of the standards and add all standards
to the calibration curve.
Steps
1 Display the Calibration Toolset.
2 Add the low-level standard to the
calibration curve.
68
Detailed Instructions
• Click the Calibration
icon, which is near
the middle of the window.
a Do one of the following:
Click the New
Calibration Table icon.
Select Calibration >
New Calibration Table.
b Click Automatic Setup Level 1.
c Click OK.
d In the Calibration Table pane
(shown below), under Compound,
type caffeine and under Amt
(amount), type 0.5.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Comments
• Do not worry at this point if your
calibration curve displays a
message that says the curve is
invalid.