4
Set Up and Run a SIM Method
Task 2. Acquire the data
Task 2. Acquire the data
Steps
1 Place the vial of sulfa sample you
prepared at 10 ng/µL into position 1
in the autosampler.
2 Inject the sulfa mix sample.
3 Monitor the total ion chromatogram
and the UV chromatogram during
data acquisition.
4 When the analysis is done, view the
results.
50
Detailed Instructions
• Click the Single Sample start
button.
a Activate the Online Plot window.
b Monitor the MS signal to ensure a
stable baseline.
a Display Data Analysis view.
b Load the data file you just created.
c Examine the DAD and MS
chromatograms.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Comments
• You prepared this sample in
"Exercise 2. Prepare the samples
for the analyses"
on page 13.
This button is present only when you
have selected Single Sample mode
from the top toolbar.
• If the baseline fluctuation for the
MS signal is greater than 10%, the
nebulizer and source chamber may
require maintenance. See the
Agilent 6100 Series Single Quad
LC/MS System Maintenance
Guide.
• If you need help, follow the general
procedure in
"Exercise 1. Display
and manipulate
chromatograms"
on page 28 in
Chapter
3.