Task 2. Enter Ms Acquisition Parameters - Agilent Technologies 6100 Series Familiarization Manual

Quadrupole lc/ms systems
Hide thumbs Also See for 6100 Series:
Table of Contents

Advertisement

2
Set Up and Run a Scan Method

Task 2. Enter MS acquisition parameters

Task 2. Enter MS acquisition parameters
Steps
1 Enter parameters for the quadrupole
mass spectrometer (MS):
Signal 1, scan mode, positive
polarity
Scan range: 100 to 500
Fragmentor:
100 V for the Agilent 6120;
125 V for the 6130 or 6150
Gain: 1.00
Threshold: 150
Stepsize: 0.10
Peakwidth: 0.05 min
Scan data storage: Condensed
Active signals: 1 only
20
Detailed Instructions
a Right-click the MSD
icon on the system
diagram and select Set
up MSD Signals.
b Enter the parameters described in
step 1
and shown in the figure
below. Take care to enter the
appropriate Fragmentor voltage for
your MS model.
c Click OK.
Set to 100 for 6120
Set to 125 for 6130 or 6150
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Comments
• To save disk space you usually
acquire line spectra (Scan Data
Storage = Condensed). However,
when you acquire spectra from
intact proteins or protein
digests/peptides, you must acquire
and deconvolute profile spectra.
(Scan Data Storage = Full).

Advertisement

Table of Contents
loading

Table of Contents