Data Acquisition in Method and Run Control View

Data Acquisition in Method and Run Control View

All method procedures and setups are described for the 1290 Infinity Binary
Pump. They can be executed in an equivalent way for the 1290 Infininty
Quaternary Pump.

Method Parameters for Test Mixture and ZORBAX RRHD Column

The 1290 Infinity LC System is supplied with a ZORBAX RRHD Eclipse Plus
C18 1.8 µm, 2.1 mm x 50 mm, column and a test mixture containing phenones
for use with this example procedure.
Phenones Test Mix (part number 5188-6529) contains 100 ng/µl each of nine
components dissolved in water / acetonitrile (65/35). The nine components
are:
• Acetanilide
• Acetophenone
• Propiophenone
• Butyrophenone
• Benzophenone
• Valerophenone
• Hexanophenone
• Heptanophenone
• Octanophenone
The method parameters for separation of this test mixture are summarized in
Table 9
96
on page 97.