Exercise 1. Set up a sequence
Exercise 1. Set up a sequence
Task 1. Prepare to create a new sequence
Steps
1 Display Method and Run Control
view.
2 Display the Sequence Toolset.
3 Display the Autosampler Tray
diagram.
4 Initiate setup of a new sequence.
5 Save the sequence under a new
name, SULFA MS SIM 1.S
52
Detailed Instructions
• In the view selection area of the
ChemStation window, click Method
and Run Control.
• In the top toolbar, click
the icon to display the
Sequence Toolset.
• Click Sampling Diagram on the
View menu.
• Select Sequence> Sequence Table.
a Select Sequence > Save Sequence
Template As.
b For Name, type
SULFA MS SIM 1.S.
c Click OK.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Comments