Exercise 1. Set Up A Sim Acquisition Method; Task 1. Load The Scan Method You Created Previously - Agilent Technologies 6100 Series Familiarization Manual

Quadrupole lc/ms systems
Hide thumbs Also See for 6100 Series:
Table of Contents

Advertisement

4
Set Up and Run a SIM Method

Exercise 1. Set up a SIM acquisition method

Exercise 1. Set up a SIM acquisition method
In this exercise, you start with your existing scan method and modify it for
SIM analysis. You keep the same LC conditions and modify only the MS
conditions. This exercise consists of the following tasks:

Task 1. Load the scan method you created previously

Steps
1 Display Method and Run Control
view.
2 Open the method
SULFA MS SCAN 1.M.
3 Save the method under a new name,
SULFA MS SIM 1.M.
44
"Task 1. Load the scan method you created previously"
"Task 2. Enter MS acquisition parameters"
Detailed Instructions
• In the view selection area of the
ChemStation window, click Method
and Run Control.
a Select File > Load > Method.
b If necessary, navigate to
C:\CHEM32\1\METHODS.
c Select SULFA MS SCAN 1.M and
click OK.
a Select File > Save As > Method.
b In the dialog box, for Name, type
SULFA MS SIM 1.M.
c Click OK.
d In the box for Comment for method
history, type a comment.
e Click OK.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
(below)
on page 45
Comments
• You save the method now to avoid
inadvertent overwrites of your scan
method.