3
Qualitative Data Analysis
Exercise 4. Print a report
Steps
2 Print the report.
3 Save the method.
42
Detailed Instructions
a Select Report > Print Report.
b After a short wait, view the Report
window.
c Verify that page 1 of the report
contains header information and an
integrated chromatogram.
d At the bottom of the report window,
click the Next button.
e Verify that page 2 of the report
shows extracted ion
chromatograms and a mass
spectrum of the first
chromatographic peak.
f Continue to click the Next button to
view results for the three additional
chromatographic peaks.
g At the bottom of the report window,
click the Print button. This prints a
hard copy of the report.
h At the bottom of the report window,
click the Close button.
a Select File > Save > Method to
overwrite the method
SULFA MS SCAN 1.M.
b In the box for Comment for method
history, type a comment.
c Click OK.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Comments
• If you wish to complete
"Set Up and Run a SIM Method",
then save the hard copy and refer
to it when you set up your SIM
groups.
• The extracted ion chromatograms
are indicators of peak purity; if the
retention times fail to coincide, the
peak likely represents more than
one compound.
• You save the method now so that
your integration parameters,
spectral display options, report
settings, and other data analysis
settings become part of the
method.
Chapter
4,