2
First Startup

Set the Checkout Method Parameters

If using an Agilent data system

Checkout method GC parameters

N O T E
58
Next, set the rest of the checkout parameters.
If using an Agilent data system, upload the current HS
parameters into the data system method. In the
ChemStation, select Method and Run Control view.
Select Instrument > Upload method from Agilent 7697A.
1
Save the method using a new name.
2
Select Instrument > Edit Parameters, then enter the GC
3
parameters listed in the next section.
Apply the method to the GC and save it.
Enter the following GC setpoints, either in the data system
(if used) or at the GC using the keypad.
The table lists the settings for a 7890A GC. Other GC types use similar
settings.
Table 11 GC checkout method
Oven
Initial temperature
Initial time
Rate
Equilibration time
Run time
Inlet (multimode, volatiles interface, or split/splitless)
Mode
Initial temperature
Pressure
Split ratio
Split flow
100 °C
4.00 min
0 (Off)
1.00 min
4.00 min
Split
250 °C (On)
134.4 kPa (19.5 psi)
20:1
80 mL/min
Installation and First Startup