Agilent Technologies 6100 Series Familiarization Manual page 75

Quadrupole lc/ms systems
Hide thumbs Also See for 6100 Series:
Table of Contents

Advertisement

Steps
4 Process the medium-level standard
and print the report.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Exercise 2. Process a sample and print a report
Detailed Instructions
a Do one of the following:
Select Report > Print Report.
Click the icon to preview
results.
b Verify that page 1 of the
report contains header information,
an integrated chromatogram, and
an external standard report.
c Check that the caffeine amount is
about 5 ng/µL.
d At the bottom of the report window,
click the Next button.
e Verify that page 2 of the report
shows the calibration curve with
the measured point identified with
dotted lines.
f (Optional) At the bottom of the
report window, click the Print
button so you get a hard copy.
g At the bottom of the report window,
click the Close button.
Quantitative Data Analysis
Comments
• Another way to generate a
hard copy is to click the
Print Report icon.