Steps
6 Refine the calibration curve.
7 (Optional) Save the method under the
same name (CAFFEINE CAL.M).
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Detailed Instructions
a Select Calibration > Calibration
Settings.
b Under Default Calibration Curve,
for Type, select Quadratic.
c Click OK.
d Verify that your calibration curve
now looks like the one below.
a Select File > Save > Method.
b In the box for Comment for method
history, type a comment.
c Click OK.
Quantitative Data Analysis
Task 5. Set up the calibration curve
Comments
6
71