Steps
4 Adjust the integration parameters to
get only four integrated peaks.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Detailed Instructions
a Click the icon for Edit/Set
Integration Events Table.
b In the Integration Events
table, for Baseline Correction,
select Advanced.
c For Height Reject, type 500000.
d Click the Integrate current
Chromatogram icon.
e Verify that your results are
similar to those shown below.
Qualitative Data Analysis
Exercise 3. Integrate the chromatogram
Comments
• For detailed information about
integration events, see Agilent
ChemStation: Understanding Your
ChemStation.
3
39