Set Up and Run a Sequence
Task 2. Edit sequence parameters
Steps
3 Enter the rest of the sequence
parameters:
54
Detailed Instructions
a Enter the following parameters
shown in the figure below.
Parts of methods to run:
According to Runtime Checklist
Wait: 10 minutes after loading a
new method
Shutdown: STANDBY
Not Ready Timeout: 15 minutes
Sequence Comment: Sequence
familiarization exercise
b Click OK.
Post-Sequence Command/Macros are
a convenient way to turn off lamps,
pumps, etc. The command or macro is
run at the end of the sequence or in
the event of an error.
Agilent 6100 Series Quadrupole LC/MS Systems Familiarization Guide
Comments
• If you wanted to run only
reprocessing (data analysis), you
would set that in Part of methods
to run.
• The Wait allows the instrument to
equilibrate when a new method is
loaded.
Two examples of Post-Sequence
Command/Macros are:
• MSSetState is a command that can
change the MS state to standby.
See the online Help for commands.
• SHUTDOWN.MAC is a macro that
will shut down the system, but you
must customize it before using it.