5
Expanding Your Measurement Experience

Testing the Agilent 8644B Internal/External 10 MHz

Testing the Agilent 8644B Internal/External 10 MHz
Defining the
Measurement
5-28
E5500 Phase Noise Measurement System Version A.02.00
This measurement example will help you measure the absolute phase
noise of an RF synthesizer.
CAUTION:
To prevent damage to the Agilent 70420A test set's hardware
components, the input signal must not be applied to the signal input connector
until the input attenuator has been correctly set for the desired configuration, as
show in
Table
5-7. Apply the input signal when the connection diagram appears.
Table 5-5
shows equipment required for this example in addition to
the phase noise test system and your unit-under-test (UUT).
NOTE: To ensure accurate measurements, you should allow the UUT
and measurement equipment to warm up at least one hour before
making the noise measurement.
Table 5-5
Required Equipment for the Agilent 8644B 10 MHz Measurement
Equipment
Agilent 8644B
Coax Cables
1
From the File menu of the E5500 User Interface, choose Open. If
necessary, choose the drive or directory where the file you want
is stored.
2
In the File Name box, choose "Conf_8644B_10MHz.pnm." See
Figure
5-23.
Quantity
Comments
1
Refer to the
"Selecting a
Reference"
on page 6-9 section of
this chapter for more information
about reference source
requirements
And adequate adapters to connect
the UUT and reference source to
the test set.
Document Part No. E5500-90024 Ed. 1.0