12
AM Noise Measurement Fundamentals
Method 1: User Entry of Phase Detector Constant
12-12
E5500 Phase Noise Measurement System Version A.02.00
Disadvantages
G
Has only moderate accuracy compared to the other calibration
methods.
Procedure
1
Using
Figure 12-9
and
all connections. If the Agilent 70420A Option 001 or Agilent
70427A is available, use one of the connection diagrams
described in
"AM Noise Measurement Block Diagrams"
page 12-3.
2
Measure the power which will be applied to the AM detector. It
must be between 0 and +23 dBm.
Figure 12-9 User Entry of Phase Detector Constant AM Noise Measurement
Setup Method 1, Example 2
Figure 12-10Modulation Sideband Calibration Setup
Figure
12-10, connect circuit and tighten
Document Part No. E5500-90024 Ed. 1.0