Table 9-3
Acceptable Amplitude Ranges for the Phase Detectors
Phase Detector
50 kHz to 1.6 GHz
Ref Input (L Port)
Signal Input (R Port)
+
15 dBm
0 dBm
to
to
+
+
23 dBm
23 dBm
* Agilent 70420A Phase Noise Test Set Options 001 and 201
Figure 9-11 Residual Connect Diagram Example
Document Part No. E5500-90024Ed. 1.0
Procedure
1
Using
Figure 9-11
and
and tighten all connections.
2
Measure the power level that will be applied to the Signal Input
port of the Agilent 70420A's phase detector.
acceptable amplitude ranges for the Agilent 70420A phase
detectors.
1.2 to 26.5 GHz
Ref Input (L Port)
+
7 dBm
to
+
10 dBm
E5500 Phase Noise Measurement System Version A.02.00
Residual Measurement Examples
Amplifier Measurement Example
Figure 9-12
as guides, connect the circuit
*
Signal Input (R Port)
0 dBm
to
+
5 dBm
9