Agilent Technologies E5500A User Manual page 280

Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

Document Part No. E5500-90024Ed. 1.0
Discriminator Measurement using FM Rate and Deviation Calibration
Figure 11-31
4
Remove the modulation from the carrier and connect your DUT.
Figure 11-32
5
The system can now run the measurement. at the appropriate
point, re-establish quadrature and continue the measurement.
Figure 11-33
The segment data will be displayed on the computer screen as the
data is taken until all segments have been taken over the entire range
you specified in the Measurement definition's Type and Range.
E5500 Phase Noise Measurement System Version A.02.00
FM Discriminator Measurement Examples
11
11-33