Document Part No. E5500-90024Ed. 1.0
3
Allow the alcohol to evaporate off the connector before making
connections
Table 20-2 Cleaning Supplies Available from Agilent Technologies
Product
Ultrajet
Alcohol wipes:
Lint-Free cloths:
Small foam swabs:
Large foam swabs
CAUTION:
Do not allow excessive alcohol to run into the connector.
Excessive alcohol entering the connector collects in pockets in the connector's
internal parts. The liquid will cause random changes in the connector's electrical
performance. If excessive alcohol gets into a connector, lay it aside to allow the
alcohol to evaporate. This may take up to three days. If you attach that connector
to another device it can take much longer for trapped alcohol to evaporate.
E5500 Phase Noise Measurement System Version A.02.00
Connector Care and Preventive Maintenance
Using, Inspecting, and Cleaning RF Connectors
Part Number
9310-6395
92193N
9310-4242
9300-1270
9300-0468
20
20-5