Chapter 1 Performance Tests
Operational Verification

10 MHz Test

1
Connect a BNC-to-BNC cable from the Reference 10MHz connector on
the instrument's rear panel to the CHANNEL 1 input connector on the
front panel.
2
Press and release the Chan Select key.
3
Verify that Ch 1 is displayed in the upper-left corner of the display.
4
Verify that the display reads 10,000,000 Hz.
5
Mark Pass or Fail in the Performance Test Record on page 1-39, Test 2.
1-8
Assembly Level Service Guide