Figure 9-1 Setup for Residual Phase Noise Measurement
Defining the
Measurement
Document Part No. E5500-90024Ed. 1.0
The setup for a residual phase noise measurement uses a phase
shifter to set quadrature at the phase detector. See
1
From the File menu, choose Open.
2
If necessary, choose the drive or directory where the file you
want is stored.
3
In the File Name box, choose "res_noise_1ghz_demoamp.pnm"
See
Figure
9-2.
Figure 9-2 Select the Parameters Definition File
4
Click the Open button.
The appropriate measurement definition parameters for this
example have been pre-stored in this file.
parameter data that has been entered for this residual phase
noise measurement example.
E5500 Phase Noise Measurement System Version A.02.00
Residual Measurement Examples
Amplifier Measurement Example
Figure
9-1.
Table 9-4
lists the
9
9-3