Defining the
Measurement
Document Part No. E5500-90024Ed. 1.0
1
From the File menu of the E5500 User Interface, choose Open.
2
If necessary, choose the drive or directory where the file you
want is stored.
3
In the File Name box, choose "StableRF.pnm". See
Figure 7-1 Select the Parameters Definition File
4
Click the Open button.
The appropriate measurement definition parameters for this
example have been pre-stored in this file.
parameter data that has been entered for the Stable RF Source
measurement example.)
NOTE: Note that the source parameters entered for step 2 in
Table 7-4
may not be appropriate for the reference source you are
using. To change these values, refer to
step 5
below. Otherwise, go to
page 7-9:
5
Using
Figure 7-2
as a guide, navigate to the Sources tab.
a
Enter the carrier (center) frequency of your UUT (5 MHz to
1.6 GHz). Enter the same frequency for the detector input
frequency.
b
Enter the VCO (Nominal) Tuning Constant (see
c
Enter the Tune Range of VCO (see
d
Enter the Center Voltage of VCO (see
e
Enter the Input Resistance of VCO (see
E5500 Phase Noise Measurement System Version A.02.00
Absolute Measurement Examples
Stable RF Oscillator
Table 7-4
Table
7-2, then continue with
"Beginning the Measurement"
Table
7-2).
Table
7-2).
Table
7
Figure
7-1.
lists the
on
Table
7-2).
7-2)
7-3