Defining the
Measurement
Document Part No. E5500-90024Ed. 1.0
FM Discriminator Measurement using Double-Sided Spur Calibration
1
From the File menu, choose Open.
2
If necessary, choose the drive or directory where the file you
want is stored.
3
In the File Name box, choose "vco_dss.pnm." See
Figure 11-2 Select the Parameters Definition File
4
Click the Open button.
The appropriate measurement definition parameters for this
example have been pre-stored in this file.
parameter data that has been entered for the FM discriminator
measurement example.
5
From the Define menu, navigate to the Measurement window.
Using
Figure 11-3
as a guide:
a
Choose the Type and Range tab from the Define
Measurement window.
b
From the Measurement Type pull-down in Type and Range
tab, select Absolute Phase Noise (using an FM discriminator).
E5500 Phase Noise Measurement System Version A.02.00
FM Discriminator Measurement Examples
Figure
Table 11-3
11
11-2.
lists the
11-5