5
Expanding Your Measurement Experience
Testing the Agilent 8663A Internal/External 10 MHz
5-18
E5500 Phase Noise Measurement System Version A.02.00
CAUTION:
The Agilent 70420A test set's signal input is subject to the
following limits and characteristics:
Table 5-3
Agilent 70420A Test Set Signal Input Limits and Characteristics
Limits
Frequency
Maximum Signal Input Power
At Attenuator Output, Operating Level
Range:
RF Phase Detectors
Microwave Phase Detectors
Internal AM Detector
Downconverters:
Agilent 70422A
Agilent 70427A
CAUTION:
To prevent damage to the Agilent 70420A Test Set's hardware components, the
input signal must not be applied to the test set's signal input connector until the
input attenuator (Option 001) has been set by the phase noise software, which will
occur at the connection diagram.
Characteristics:
Input Impedance
AM Noise
50 kHz to 26.5 GHz
+30 dBm
0 to +23 dBm
0 to +5 dBm
0 to +20 dBm
0 to +30 dBm
+5 to +15 dBm
50 ohm Nominal
dc coupled to 50 ohm load
Document Part No. E5500-90024 Ed. 1.0