Method 3: Single-Sided-Spur

Document Part No. E5500-90024Ed. 1.0
Advantages
G
Will measure source without modulation capability.
G
Calibration is done under actual measurement conditions so all
non-linearities and harmonics of the AM detector are calibrated
out. The double-sided spur method and the single-sided-spur
method are the two most accurate methods for this reason.
Disadvantages
G
Requires 2 RF sources, which must be between 10 Hz and 40 MHz
apart in frequency.
G
Requires an RF spectrum analyzer for manual measurement of
the signal-to-spur ratio and spur offset.
Procedure
1
Connect circuit as shown in
connections. If the Agilent 70420A Option 001 or Agilent 70427A
is available, use one of the connection diagrams described in
Noise Measurement Block Diagrams"
Figure 12-18AM Noise Measurement Setup Using Single-Sided-Spur
2
Measure the power which will be applied to the AM detector. It
must be between 0 and +23 dBm.
E5500 Phase Noise Measurement System Version A.02.00
AM Noise Measurement Fundamentals
Method 3: Single-Sided-Spur
Figure
12-18, and tighten all
on page 12-3.
12
"AM
12-19