Synthesized Residual Measurement Using Beatnote Cal - Agilent Technologies E5500A User Manual

Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

8
Residual Measurement Fundamentals
Calibration Options
Synthesized
Residual
Measurement using
Beatnote Cal
Procedure
Figure 8-13 Synthesized Residual Measurement using Beatnote Cal
8-18
E5500 Phase Noise Measurement System Version A.02.00
This calibration option requires two synthesizers of which one must
be tunable such that a beatnote can be acquired. For the system to
calibrate, the beatnote frequency must be within the following ranges
shown in
Table
8-5.
Table 8-5
Frequency Ranges
Carrier Frequency
<500 kHz
<5 MHz
<50 MHz
<250 MHz
>250 MHz
or 1/2 the frequency range of the configured analyzer, or whichever is lower.
1
Connect circuit as per
2
Offset the carrier frequency of one synthesizer to produce a
beatnote for cal.
3
After the phase noise system reads the beatnote, set the software
to the same carrier frequency.
Beatnote Frequency Range
10 Hz to 10 kHz
10 Hz to 100 kHz
10 Hz to 1 MHz
10 Hz to 10 MHz
10 Hz to 50 MHz
Figure
8-13, and tighten all connections.
Document Part No. E5500-90024 Ed. 1.0