Testing the Agilent 8663A Internal/External 10 MHz
Document Part No. E5500-90024Ed. 1.0
This measurement example helps you measure the absolute phase
noise of an RF synthesizer.
CAUTION:
To prevent damage to the Agilent 70420A test set's hardware
components, the input signal must not be applied to the signal input connector
until the input attenuator has been correctly set for the desired configuration, as
shown in
Table
5-3. Apply the input signal when the connection diagram
appears.
Table 5-1
shows equipment required for this example in addition to
the phase noise test system and your unit-under-test (UUT).
NOTE: To ensure accurate measurements, you should allow the UUT
and measurement equipment to warm up at least one hour before
making the noise measurement.
Table 5-1
Required Equipment for the Agilent 8663A 10 MHz Measurement
Equipment
Agilent 8663A
Coax Cables
E5500 Phase Noise Measurement System Version A.02.00
Expanding Your Measurement Experience
Testing the Agilent 8663A Internal/External 10 MHz
Quantity
Comments
1
Refer to the
Reference"
this chapter for more information
about reference source
requirements
And adequate adapters to connect
the UUT and reference source to
the test set.
5
"Selecting a
on page 6-9 section of
5-11