Baseband Noise Measurement Examples

Baseband Noise without using a Test Set Measurement Example

Baseband Noise without using a Test Set Measurement Example
Defining the
Measurement
14-6
E5500 Phase Noise Measurement System Version A.02.00
This measurement example will help you measure the noise voltage
of a source.
NOTE: To ensure accurate measurements, you should allow the UUT
and measurement equipment to warm up at least one hour before
making the noise measurement.
1
From the File menu, choose Open.
2
If necessary, choose the drive or directory where the file you
want is stored.
3
In the File Name box, choose
"BBnoise_without_testset_89410.pnm."
Figure 14-5 Select the Parameters Definition File
4
Choose the OK button. The appropriate measurement definition
parameters for this example have been pre-stored in this file.
Table 14-2
lists the parameter data that has been entered for this
measurement example.
Document Part No. E5500-90024 Ed. 1.0