Bent Semirigid
Cables
Other Multipin
Connectors
Document Part No. E5500-90024Ed. 1.0
G
Carefully insert the male connector center pin into the female
connector. (Try to make sure the cable is aligned with the
instrument connector properly before joining them.)
G
Turn the silver nut clockwise by hand until it is snug, then tighten
with an 8 inch-lb torque wrench (part number 8720-1765). This
wrench may be ordered from Agilent Technologies.
Semirigid cables are not intended to be bent outside of the factory.
An accidental bend that is slight or gradual may be straightened
carefully by hand. Semirigid cables that are crimped will affect
system performance, and must be replaced. Do not attempt to
straighten a crimped semirigid cable, its performance will not be
restored.
There are other multipin connectors in the system (Agilent MSIB, for
example). These are sometimes held in place by a pair of screws.
E5500 Phase Noise Measurement System Version A.02.00
Connector Care and Preventive Maintenance
Removing and Reinstalling Instruments