13
AM Noise Measurement Examples
AM Noise using an Agilent 70420A Option 001
Figure 13-3 Navigate to AM Noise
13-4
E5500 Phase Noise Measurement System Version A.02.00
NOTE: The amplitude of a source under system control, for an AM
noise measurement, will automatically be set to +10 dBm. If any
other amplitude is desired, the source should be placed under
manual control. All other measurements set the source to +16 dBm
automatically.
The appropriate measurement definition parameters for this
example have been pre-stored in this file.
parameter data that has been entered for the FM Discriminator
measurement example.)
5
From the Define menu, choose Measurement; then choose the
Type and Range tab from the Define Measurement window.
6
From the Measurement Type pull-down, select AM Noise. See
Figure
13-3.
7
Choose the Sources tab from the Define Measurement window.
Table 13-3
lists the
Document Part No. E5500-90024 Ed. 1.0