Mms Module Removal And Reinstallation - Agilent Technologies E5500A User Manual

Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

MMS Module
Removal and
Reinstallation
Document Part No. E5500-90024Ed. 1.0
To Remove an MMS Module
1
Set the mainframe line switch to OFF.
2
Remove all rear panel cables going to the module
3
On the bottom of the mainframe front panel is a dark-colored,
horizontal access panel. Pry outward at the top of this panel to
open it.
4
With an 8 mm hex-ball driver, loosen the module hex nut latch.
5
Go to the back of the system and press against the module's rear
panel and slide the module forward several inches.
6
From the front of the system, pull the module out.
To Reinstall an MMS Module
1
Set the MMS mainframe line switch to OFF.
2
Check the GPIB address switch on the module for the correct
address setting. Refer to the manual for the MMS module for
information on the Agilent MSIB switch. For proper address
settings, refer to the system information chapter.
CAUTION:
Reinstalling an MMS module without setting the GPIB address
will cause the system to malfunction. (Not all MMS modules use GPIB settings.)
3
On the bottom of the mainframe front panel is a dark-colored,
horizontal access panel. If necessary, pry outward at the top of
this panel to open it.
4
Slide the module into the mainframe.
5
Press against the front of the module while tightening the hex-nut
latch (with an 8 mm hex-ball driver).
6
Close the access panel.
7
Go to the back of the system and connect intermodule cables.
E5500 Phase Noise Measurement System Version A.02.00
Connector Care and Preventive Maintenance
Removing and Reinstalling Instruments