Agilent 8643A Frequency Limits - Agilent Technologies E5500A User Manual

Hide thumbs Also See for E5500A:
Table of Contents

Advertisement

Agilent 8643A Frequency Limits

Table 17-2 Agilent 8643A Frequency Limits
Note: Special Function 120 must be enabled for DCFM
Model
Option
Band Minimum
Number
(MHz)
8643A
002
1030
8643A
002
515
8643A
Standard
515
8643A
Both
257.5
8643A
Both
128.75
8643A
Both
64.375
8643A
Both
32.1875
8643A
Both
16.09375
8643A
Both
8.046875
8643A
Both
4.0234375
8643A
Both
2.01171875
8643A
Both
1.005859375
8643A
Both
0.5029296875
8643A
Both
0.25146484375
1
Takes into account limited tuning resolution available in linear FM (Special Function 120, refer to
Special Functions"
on page 17-16).
2
The Agilent 8643A defaults to Mode 2 operation.
3
Wideband FM: Use Special Function 125 (refer to
Agilent 8643A Mode
Keys
Document Part No. E5500-90024Ed. 1.0
Band Maximum
(MHz)
2060
1029.99999999
1030
514.99999999
257.49999999
128.74999999
64.37499999
32.18749999
16.09374999
8.04687499
4.02343749
2.01171874
1.005859365
0.5029296775
"How to Access Special Functions"
G
The [Mode 1] key provides the maximum FM deviation and
minimum RF output switching time. Noise level is highest in this
mode, as shown in the following table.
G
The [Mode 2] key provides a median range of FM deviation and
RF output switching time, as shown in
Agilent 8643A defaults to Mode 2 operation.
E5500 Phase Noise Measurement System Version A.02.00
Reference Graphs and Tables
Agilent 8643A Frequency Limits
1
Minimum Recommended PTR (Peak Tune Range)
PTR =FM Deviation x VTR
2
Mode 2
Mode 1
2000000
20000000
1000000
10000000
1000000
10000000
500000
5000000
250000
2500000
125000
1250000
62500
625000
31200
312000
15600
156000
7810
78100
3900
39000
1950
19500
976
9760
488
4880
on page 17-16).