Setting Up And Operating The Injector System (Optional); Setting Up The Injector System - Promega GloMax Operating Manual

Discover system
Hide thumbs Also See for GloMax:
Table of Contents

Advertisement

4.

Setting Up and Operating the Injector System (optional)

4.A. Setting Up the Injector System
Components
Unpack the GloMax
component list is enclosed with the Injector System for details. The Injector System components are shown in
Figure 87.
Figure 87. GloMax
System Console with Vertical Support Rods, Inlet Tubing Assembly and Inlet Tube Holder. The DB-15 Data
Cable, Waste Collection Tray and Outlet Tubing Assembly are also shown.
Injector System Installation
1.
Turn off the GloMax
System. Failure to turn off the instrument
!
may cause damage to the instrument
electronics.
2.
Place the Injector System on top of the in-
strument with the four rubber feet within the
indentation on the instrument top, and make
sure that the Injector System is level.
78
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
TM397 · Revised 8/18
Discover Injector System. Verify all components and accessories have been received. A
®
Discover Injector System with accessories. Shown assembled here are the Injector
®
Discover
®
Figure 88. Positioning the Injector System on
top of the GloMax
Discover System.
®
www.promega.com