Self Test
Each time the instrument is powered "ON", it automatically performs a self test to check the motions of all axes.
This button allows users to manually initiate a self test.
Tools
Tools may be accessed from the Tools icon in the menu bar (Figure 62, Panel A) or using the Tools button on the
'Settings' screen (Figure 62, Panel B).
A.
Figure 62. Accessing Tools. Panel A. Accessing the tools from the menu bar. Panel B. 'Tools' screen accessed
by pressing the Tools button on the 'Settings' screen.
Heater
Enter the desired temperature in °C between 25–45°C. The heater will maintain a temperature within 1°C of the
set point.
!
Note: The instrument is fan-cooled. The temperature cannot be set lower than the ambient temperature of your
laboratory.
Figure 63. Heater Symbol.
The Heater will remain on after the run unless manually turned OFF by the user or the PC is disconnected from
!
the instrument.