Promega GloMax Operating Manual page 80

Discover system
Hide thumbs Also See for GloMax:
Table of Contents

Advertisement

3.
Remove the front panel from the pump
station to reveal tubing attachment points.
4.
Attach the fitting at the end of the outlet
tubing assembly to the inside port of the
injector pump. Ensure that the line to
Injector #1 (left) goes to the inside port of
the left pump and the line for Injector #2
(right) goes to the inside port of the right
pump.
5.
On the injector system console, insert the
thicker end of the vertical support rod into
the small opening. Press down firmly until
the rod clicks into place.
6.
Slide the inlet tube holder onto the vertical
support rod.
7.
Attach the fitting at the end of the inlet
tubing to the outside port of the injector
pump. Ensure that the injector tubing
attaches to the correct port. Refer to the
diagram located in the front of the pump for
details.
8.
Pinch the lever on the back of the inlet tube
holder to clip the stainless steel tube securely
into place.
9.
Screw the injector fittings into the top of the
GloMax
®
Note: The tubing inside the instrument is
installed at the factory.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
www.promega.com
Discover Instrument.
Figure 89. Attaching the fitting to the inside
port.
Figure 90. Inserting the vertical support rod
and inlet tube holder.
Figure 91. Injector showing attachment of
inlet and outlet tubing fittings.
Figure 92. Installation of the inlet tube holder
clip.
79
TM397 · Revised 8/18