3.C. Protocol Screen (continued)
Heat Function
Heat the plate by using the Heat function. To prewarm the heater prior to the measurement read, select Heat
from the Tools icon of the menu bar, and select Wait in the Heating step. If Wait is selected in the protocol, the
measurement read will begin only when the temperature has reached the set point. If Wait is not selected, the
protocol will proceed whether or not the heater has reached the set point.
Enter the desired temperature, in degrees Celsius, and between 25 and 45°C. The heater will maintain
temperature within 1°C of the set point.
!
Note: The instrument is fan cooled; the temperature cannot be set lower than the ambient laboratory temperature.
Figure 25. The Heat Function.
The Heater may also be accessed from the Tool icon in the menu bar.
A.
Figure 26. Accessing the Heater from the Tools icon in the menu bar.
!
Note: The heater will continue to remain ON after the run unless it is turned OFF or the PC is disconnected from
the instrument.