2.B. Switching On the GloMax
Exit
F
igure 6. GloMax
2.C. Shutting Down the GloMax
1.
Switch the instrument OFF using the On/Off switch on the back of the instrument.
Note: When the instrument is not in use for a prolonged period, it is best to turn OFF the external power
supply using its own On/Off switch.
2.
Shut down the GloMax
3.
Shut down the Tablet PC using the Windows
Windows 8: Swipe in from the right edge of the screen, select Settings, then select Power: Shut Down.
Windows 10: From the Desktop screen, swipe upward to expose the Taskbar. Touch the Start button to
open the Start menu, then touch the Power icon on the left side of the Start menu. From the Power drop-
down list, touch the Shut Down item.
!
Note: When not in use for a prolonged period of time, the Tablet PC should be powered OFF.
2.D. Operating the GloMax
Open the Door of the instrument by tapping the Door icon on the top right corner of the screen (Figure 6).
1.
When the plate tray extends, add the plate such that well A1 of the plate corresponds to the label "A1" on
the plate tray (Figure 7).
2.
Select one of the yellow Quick Read protocols on the 'Home' screen, or access a protocol of interest by
pressing the Protocols button.
14
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
TM397 · Revised 8/18
Discover Instrument (continued)
®
Back
Discover software 'Home' screen.
®
Discover Instrument
®
Discover Software by tapping on the x in the upper left corner (Figure 6).
®
Discover System
®
shutdown procedure appropriate for your tablet PC:
®