Promega GloMax Operating Manual page 5

Discover system
Hide thumbs Also See for GloMax:
Table of Contents

Advertisement

1.B. Specifications (continued)
Luminescence Module Specifications
Detector
Wavelength Range
Detection Limit
Linear Dynamic Range
Cross Talk
Filtered Luminescence
Fluorescence Module Specifications
Detector
Light Source
Read Position
Wavelength Selection
Wavelengths Included
Detection Limit
Linear Dynamic Range
4
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
TM397 · Revised 8/18
Top-reading, head-on photon-counting photomultiplier tube (PMT)
350–700nm
3 × 10
moles of luciferase
–21
9 logs
Less than 3 × 10
(white, 96-well plate, Corning Cat.# 3912)
–5
Filter paddle containing 5 standard filters for luminescence
including: 495nm SP, 530nm LP, 540nm SP, 600nm LP and 450nm
(8nm BP). Additional custom emission filters may also be used.
Contact Promega Technical Services.
PIN photodiode
Wavelength-matched LED
Top-reading
Filter paddle with 5 standard excitation and
emission filters. Empty filter positions are available
for custom excitation and emission wavelengths.
Contact Promega Technical Services.
UV (Ex: 365nm, Em: 415–445nm)
Blue (Ex: 475nm, Em: 500–550nm)
Green (Ex: 520nm, Em: 580–640nm)
Red (Ex: 627nm, Em: 660–720nm)
AFC (Ex: 405nm, Em: 495–505nm)
2fmol fluorescein/200µl
6 logs (assay-dependent)
www.promega.com