Promega GloMax Operating Manual page 83

Discover system
Hide thumbs Also See for GloMax:
Table of Contents

Advertisement

4.B. Operating the Injector System (continued)
Priming Injectors
1.
Place the reagent container (tube or bottle) into the reagent holder.
2.
Cover the bottle opening with Parafilm
3.
Insert the stainless steel tube through the Parafilm
4.
Clip the stainless steel tube into the slot on the horizontal bar, and ensure the tube reaches the bottom of the
reagent bottle.
Tap the Prime button shown in Figure 95. The door will open.
5.
Figure 95. Injector prime function.
6.
Place the Waste Collection Tray on the Microplate Sample Tray.
7.
Indicate the injector to prime. Set the volume.
Note: The minimum volume needed to prime is 500µl.
Tap the Start Prime button
8.
9.
After priming is complete, remove the Waste Collection Tray.
Once the injectors are primed, they are ready for use according to the programmed protocol.
82
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
TM397 · Revised 8/18
to prevent reagent evaporation.
®
and into the reagent bottle.
®
www.promega.com