3.G. 'Settings' Screen (continued)
Injectors
Dual Injectors are an optional accessory for the GloMax
Injector System Specifications
Number of Injectors
Dispense Volume Range
Plate Compatibility
Injection Speed
Waste Collection Tray Volume
Void Volume
*For standard Luciferase Assay or Dual-Luciferase
Warning: Make sure sufficient volume is available in the well to accept the injected volume, especially when
!
injectors are used within a loop function. Failure to do so may cause an overflow of reagent and/or sample into
adjacent wells or into the interior of the instrument.
The Injectors may also be accessed from the Tools icon in the menu bar.
See Setting Up and Operating the Injector System (Section 4) for more details.
Filters
1.
To change filter slides for BRET/filtered luminescence or for fluorescence (including FRET), select the
appropriate Reconfigure... button.
Figure 64. Initial screen showing instrument filter configuration.
64
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
TM397 · Revised 8/18
Discover Instrument.
®
Two Injectors
5–200µl in 1µl increments
6-, 12-, 24-, 48- and 96-well plates
20–500µl/second*
Approximately 50ml
500µl
Assays, we recommend an injector speed of 200µl per second.
®
www.promega.com