O P E R A T I N G M A N U A L
Maxwell
®
Operating Manual
Instructions for use of the Maxwell® RSC Instrument,
Catalog Number AS4500.
RSC Instrument
Revised 4/15
Revised 4/15
TM411
TM411

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Maxwell RSC Instrument and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Promega Maxwell RSC Instrument

  • Page 1 O P E R A T I N G M A N U A L Maxwell RSC Instrument ® Operating Manual Instructions for use of the Maxwell® RSC Instrument, Catalog Number AS4500. Revised 4/15 Revised 4/15 TM411 TM411...
  • Page 2: Table Of Contents

    5.E. Reports ........................54 5.F. Running Reports ...................... 57 5.G. Sanitizing ......................... 58 Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 3 10. Certificate of Decontamination ................... 72 11. Summary of Changes ......................73 Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 4: Introduction

    The Maxwell RSC is for research use only. ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 5: Product Use Limitations

    16mm, 4W, 0.17A current, 29V, spectral peak F 253.7, UV output 0.9W Installation Category II Pollution Degree 2 Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 6: Product Components

    PC, Tablet PC Power Cords, Tablet PC Holder and Quantus™ Fluorometer for the Maxwell RSC Instrument. ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 7: Precautions

    Do not load any additional software programs on the Tablet PC supplied with Maxwell ® Instrument. Additional programs may cause the application to slow down. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 8: Safety Symbols And Marking

    Serial Number Voltage: 95-240 VAC Made in 50-60Hz, 1.0A Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 250V, 2.5A TM411 · Revised 4/15 www.promega.com Time-lag, 5x20mm 249469 050-9800 Rev.
  • Page 9 CE Compliance Marked FCC Compliance Marked KCC Compliance Marked Lot Number Important Information Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 10: Environmental Requirements (Operating, Shipping And Storage Conditions)

    PC into an uninterruptible power supply. Hardware Overview Figure 2. Front of the Maxwell RSC Instrument. ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 11 RSC Instrument. The magnet and plunger bars, which are used for sample processing, and the platform contain- ing the deck tray are evident. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 12: Unpacking The Maxwell ® Rsc Instrument

    Save the packaging material in case the instrument needs to be returned for service or repair later. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 ·...
  • Page 13: Setting Up The Maxwell ® Rsc Instrument

    RSC Instrument door is spring-loaded and will close itself once the foam pieces are ® removed from the inside of the instrument. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 14: Setting Up The Tablet Pc

    Figure 9. Tablet PC Holder on top of the Maxwell RSC Instrument. ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 15 RSC Power Cable to the back of the Maxwell RSC Instrument. ® ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 16: Switching On The Maxwell ® Rsc Instrument

    The deck, plunger bar and magnetic rod assembly are moved to check operation. On/Off switch Figure 13. On/Off Switch. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 17: Shutting Down The Maxwell ® Rsc Instrument

    When not in use for a prolonged period of time, the Tablet PC should be unplugged. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 18: Configuring The Tablet Pc

    11. Close the Control Panel by touching the red X button in the upper right corner of the screen. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 ·...
  • Page 19 Note: If the Local Users and Groups menu item is not visible, select the System Tools item to expand that menu heading. Figure 15. Selecting Groups from Computer Management Screen. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 20 PromegaAdministrators group. Close the Computer Management window. Figure 16. Adding users to the PromegaAdministrators group. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 21 PromegaAdministrators group. Close the Computer Management window. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 22 Figure 19. Instrument Name Window. Administrators can manually enter the instrument name from the Instrument Name window. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 23: Maxwell ® Rsc User Interface Screens

    Help. Pressing the help button will activate the context-sensitive help for the current page of the • Maxwell RSC User Interface. ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 24 From the Maxwell RSC Administrator page (Figure 23) touch the Sample Entry button. ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 25 RSC Settings ® ® screen from which Administrator options can be accessed. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 26 RSC Administrator page. The Administrator page is only visible to administrators ® and gives access to sample entry and additional options. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 27 The administrator can choose whether any of the available sample-specific fields are required to match the Sample ID for each processing position on the instrument. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 28 (in minutes) and other settings such as Sanitize after extraction and/or Sanitize on instrument startup for a specified amount of time. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 29 ® provide instructions on how to update the Maxwell RSC Software. ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 30 (Figure 28). This file will be used to import a new or updated method. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 ·...
  • Page 31 By default all options are unchecked. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 32 ® Maxwell RSC Instrument. ® Figure 31. Self Test protocol being executed. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 33 RSC Settings screen. From this screen you can perform “Clean Up” for the ® Maxwell RSC Instrument. ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 34 RSC Instrument is calibrated at the manufacturing site prior to use. No user calibration ® is necessary. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 35: Operating The Maxwell ® Rsc Instrument

    RSC Home screen. Selecting the Start button begins the process of running ® a protocol on the Maxwell RSC Instrument. ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 36 Figure 35. Kit label indicating bar code to scan. Shown in the blue box is the bar code to scan on the kit label for starting a purification run. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 37 Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 38 • Tap twice on the Sample ID box to enter the sample information manually. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 39 To select/deselect a cartridge position touch the blue rectangle for any given position. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 40 Figure 40. Door open screen. Informs the user that the Maxwell RSC Instrument door will open. ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 41 Extraction Checklist and should be performed prior to this stage according to the technical manual for the kit being processed. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 42 Place the deck tray in the instrument as shown in Figure 43. Ensure that the deck tray is completely seated inside the instrument. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 43 Abort button in the lower right corner of the screen. Note: Any samples being processed will be lost if a run is aborted. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 44 If you wish to abort the current run, touch the Abort button in the lower right corner of the screen. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 45 Important: Used cartridges and plungers should be disposed of appropriately according to your institution’s procedures for hazardous and biohazardous waste. Do not reuse reagent cartridges, plungers or elution tubes. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 46 UV sanitization (see Section 5.G). Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 ·...
  • Page 47 An aborted run (initiated by the user or due to instrument error) will result in the loss of all the samples. Do not attempt to repurify samples from an aborted run. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 48 An aborted run (initiated by the user or due to instrument error) will result in the loss of all the samples. Do not attempt to repurify samples from an aborted run. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 49: Clean Up

    Report View window. If the plunger Clean Up fails, you should contact Promega Technical Services for further assistance. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 50: Quantitating With The Quantus™ Fluorometer

    Quantus™ Fluorometer needs calibration or the mode needs to be changed, consult the Quantus™ Fluorometer Operating Manual #TM396. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 51 Open the Quantus™ Fluorometer lid by pressing down on the lower right corner of the sample lid. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 52 10. To move to the next sample, either press the arrow buttons on the lower right or left of the Quantus™ Fluorometer Measurement screen or tap the sample number. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 53 15. The quantitation data will be stored, and the run report will be displayed. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 54: Reports

    In the lower left corner of the screen is an Export button that will be used for exporting report information. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 55 Quantus™ Fluorometer is plugged into one of the USB ports on the back of the Maxwell ® RSC Instrument. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 56 (i.e., LIMS). An example of the tab-separated format file is shown in Figure 59. Figure 59. Example of an exported text file in the tab-separated format. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 57: Running Reports

    Selecting a Date button on the left side of the window will filter the reports displayed based on their date. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 58: Sanitizing

    Figure 63. Home screen. Selecting the Sanitize button begins the process of sanitizing the Maxwell ® RSC Instrument. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 59 Figure 65. Example Sanitization Report. The sanitization report will be displayed after completion of sanitization. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 60: Cleaning And Maintenance

    If the magnetic rod assembly cannot be cleaned, please contact Promega Technical Services for assistance. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 61: Dealing With Spills

    Bleach should not be used to clean Maxwell RSC reagent spills. ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 62: Troubleshooting

    Do not attempt to repurify samples from an aborted run. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 63: Errors And Warnings

    User warning informing user that a firmware version detected change has been detected. Contact Promega Technical Services. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 64: Use Of A Usb Flash Drive

    Note: If the original packaging is lost or damaged, contact Promega or your local Promega representative for replacement packaging. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 65: Repacking The Maxwell ® Rsc Instrument

    This is by design. Do not try to force the door to close, or the door can break. Place the instrument back in the plastic bag. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 66 Place the top foam packaging over the top of the instrument as shown (Figure 68). Figure 68. Foam packing on top of box. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 67: Instrument Disposal

    Instrument must be decontaminated prior to disposal. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...
  • Page 68: Warranties, Service Agreements And Related Products

    Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 69 Train customer(s) to use the log book • Complete customer-specific log book, instrument sticker and OQ documentation Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 70 Contact Promega or your local Promega representative. We will be happy to assist you by telephone at no charge. Repair service will be quoted prior to any work being performed. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15...
  • Page 71: Related Products

    1 each AS3200 ® Maxwell RSC/CSC Deck Tray 1 each SP6019 ® Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com TM411 · Revised 4/15...
  • Page 72: Certificate Of Decontamination

    This instrument is certified by the undersigned to be free of radioactive contamination. Date: Place: Signature: Name (block capital letters): Title: Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM411 · Revised 4/15 www.promega.com...
  • Page 73: Summary Of Changes

    Product claims are subject to change. Please contact Promega Technical Services or access the Promega online catalog for the most up-to-date information on Promega products. Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 www.promega.com...