Promega Maxwell CSC 48 Operating Manual

Promega Maxwell CSC 48 Operating Manual

Instrument ivd mode

Advertisement

Quick Links

Maxwell
CSC 48 Instrument IVD Mode
®
Operating Manual
Instructions for use of the Maxwell
CSC 48 Instrument when running IVD Mode,
®
Model Number AS8000
For use in RUO Mode, see the Maxwell
CSC 48 RUO Mode Operating Manual #TM628.
®
Ensure that all sealing tape and any residual adhesive are removed from the Maxwell
CSC reagent
®
cartridges before placing the cartridges into the instrument.
instructions for
use of product
AS8000
TM623 | Revised 6/20

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Maxwell CSC 48 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Promega Maxwell CSC 48

  • Page 1 Maxwell CSC 48 Instrument IVD Mode ® Operating Manual Instructions for use of the Maxwell CSC 48 Instrument when running IVD Mode, ® Model Number AS8000 For use in RUO Mode, see the Maxwell CSC 48 RUO Mode Operating Manual #TM628. ®...
  • Page 2: Table Of Contents

    (Operating, Shipping and Storage Conditions) ........11 2 Hardware Overview ................12 3 Unpacking the Maxwell CSC 48 Instrument ........14 ® 3.1 Setting Up the Maxwell CSC 48 Instrument ..........15 ® Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 3 7 Cleaning and Maintenance ...............66 7.1 General Care ..................66 7.2 Cleaning the Hardware ................67 7.3 Dealing with Spills ...................68 8 Troubleshooting ................70 8.1 Use of a USB Flash Drive ................72 Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 4 All technical literature is available at: www.promega.com/protocols/ Visit the Web site to verify that you are using the most current version of this Technical Manual. E-mail Promega Technical Services if you have questions on use of this system: techserv@promega.com Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 5: Introduction

    Place the deck trays containing prepared cartridges into the instrument, and the method will run automatically. Using the instrument does not require any special training; however, training is available as part of Operational Qualification (OQ; see Section 10.2). Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 6: Product Intended Use

    • UV lamp to aid in decontamination of instrument • Integrated Vision system for confirming proper deck tray preparation • Integral USB hub for easy connection to accessory devices • Comprehensive technical support Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 7: Maxwell ® Csc 48 Instrument Specifications

    • 1 UV Bulb (installed) • 1 Bar Code Reader • 1 Tablet Mount Attachment Pieces • 1 2.5mm Hex Wrench • 1 Stylus • 1 Quick Start Guide • 1 Setup Guide Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 8: Inspection

    CSC 48 Instrument, please inspect the package carefully to make ® sure all components are present and that the instrument has not been damaged in shipping. If any item is damaged, contact Promega Technical Services (e-mail: techserv@promega.com). Included components are shown in Figure 1. Figure 1. Maxwell CSC 48 Instrument.
  • Page 9: Precautions

    • Use only Promega-supplied cartridges, plungers and elution tubes. • Do not reuse cartridges, plungers or elution tubes. • If the equipment is used in a manner other than that specified by Promega, the protection provided by the equipment may be impaired.
  • Page 10: Safety Symbols And Marking

    Avertissement. Risque biologique. Avertissement. Risque de lampe Warning. UV light hazard. Do not look UV. Ne regardez pas directement en directly at the UV light. direction de la lampe UV. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 11 95-240 VAC It is important to understand and follow all laws regarding the Made in z, 1.0A safe and proper disposal of electrical instrumentation. Please consult your local Promega representative regarding instrument 0V, 2.5A disposal. g, 5x20mm 249469 050-9800 Rev. D...
  • Page 12: Environmental Requirements (Operating, Shipping And Storage Conditions)

    CSC 48 Instrument and Tablet PC to an uninterrupted power supply ® (UPS). This is to ensure continued operation during brief power outages that otherwise might result in aborted instrument runs and lost samples. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 13: Hardware Overview

    On/Off switch and the 3-port USB hub for accessories such as the Bar Code Reader. Panel B. The rear of the instrument showing the power cable connector and USB communication port for the Tablet PC. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 14 Figure 4. Magnetic assembly components and platform. The hardware components inside the Maxwell CSC 48 Instrument. The magnet and plunger bars, which are used for sample ® processing, and the platform, which holds the deck trays, are evident. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 15: Unpacking The Maxwell ® Csc 48 Instrument

    5. Place the instrument on a flat stable surface. Leave at least 7.5 inches (19cm) of clearance in front of the instrument to allow the instrument door to open without hindrance. 6. Remove the plastic bag from around the instrument. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 16: Setting Up The Maxwell ® Csc 48 Instrument

    Panel B. Remove the four shipping screws from the sides of the front magnet and plunger bars. Figure 8. Removing the top foam from the instrument. Move the magnet and plunger bars downward to remove the top foam piece. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 17: Preparing The Maxwell ® Csc 48 Instrument For Use

    Slide the Tablet PC into the tablet holder. Secure the Tablet PC in the tablet holder by using the 2.5mm Hex Wrench to screw the Tablet Mount Attachment Piece into place at the top of the tablet holder. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 18 Instrument. Connect an ethernet cable (not included) to the USB Ethernet Adapter and an ethernet port. Figure 11. Maxwell CSC 48 Instrument with the Tablet PC. ® 7. The Maxwell CSC 48 Instrument is now ready to use. ® Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 19: Switching On The Maxwell ® Csc 48 Instrument

    Figure 12. Instrument power switch and power button. Panel A. Rocker On/Off power switch on the back of the instrument. Panel B. Power button on the right side of the instrument. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 20: Shutting Down The Maxwell ® Csc 48 Instrument

    Instructions for configuring the Tablet PC can be found in the Maxwell ® should configure the CSC Tablet PC Configuration Manual #TM484. Tablet PC according to the IT rules and IT procedures pertinent to your site. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 21: Maxwell ® Csc 48 User Interface

    Administrator (only available to users with Administrator level access to the Maxwell CSC 48 software). See Section 5.2, User Interface ® Settings, and Section 5.3, Administrator Settings. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 22 This icon toggles the open/closed state of the door on the Maxwell ® Door CSC 48 Instrument. Pressing this icon will activate the context-sensitive help for the current Help screen of the Maxwell CSC 48 User Interface. ® Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 23: User Interface Settings

    CSC 48 ‘Settings’ screen. From this screen various instrument-specific ® functions can be accessed. The Administrator button will be displayed only for Windows user ® accounts with administrator-level access to the Maxwell CSC 48 software. ® Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 24 • Plunger Calibration Value: the calibration value for the plunger bar axis on the Maxwell CSC 48 Instrument ® • Magnet Calibration Value: the calibration value for the magnet bar axis on the Maxwell CSC 48 Instrument ® Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 25 Maxwell® CSC 48 User Interface 5.2 User Interface Settings (continued) Figure 15. ‘About Maxwell CSC 48’ screen. Instrument- and software-specific information is ® displayed on the ‘About Maxwell CSC 48’ screen. ® Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 26 Following the Self Test, the software automatically opens this report. Figure 16. Maxwell CSC 48 ‘Settings’ screen. From this screen you can perform a ® Self Test for the Maxwell CSC 48 Instrument. ® Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 27 Clean Up process and will generate and display the system report from the Clean Up process. If plungers are not ejected after several clean up attempts, the operator should contact Promega Technical Services (techserv@promega.com; 1-800-356-9526) to determine the next appropriate steps.
  • Page 28 Figure 18. ‘Clean Up Checklist’ screen. This screen presents you with a checklist of items that must be performed and confirmed before attempting to unload plungers. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 29 Instrument logs will be exported as a zip file to the path that was specified by the operator. The zip file of the logs should be retrieved from the specified location and sent to a Promega Technical Services representative for further troubleshooting.
  • Page 30: Administrator Settings

    ‘Administrator Page’ screen when starting from the ‘Home’ screen in the software. Figure 20. Maxwell CSC 48 ‘Home’ screen. Selecting Settings opens the Maxwell ® ® 48 ‘Settings’ screen. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 31 Figure 21. Maxwell CSC 48 ‘Settings’ screen. The Administrator button will only be visible if the ® currently logged in operator has administrator-level rights within the Maxwell CSC 48 software. ® Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 32 CSC 48 ‘Administrator Page’ screen. The ‘Administrator Page’ screen ® is only visible to administrators and gives access to customize the behavior of several aspects of the Maxwell CSC 48 system. ® Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 33 "Must Match Sample ID" text for that field. Once enabled the box will contain a check mark. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20...
  • Page 34 It is possible to specify the duration of the UV treatment and when UV light treatment of the instrument should be performed. 1. From the ‘Administrator Page’ screen, touch the Sanitization Settings button to open the ‘Sanitization Settings’ screen. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 35 To discard any changes to the UV sanitization options, touch the Cancel button. Touching either button will result in a return to the ‘Administrator Page’ screen. Figure 24. The ‘Sanitization Settings’ screen. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 36 2. Touch the Back icon in the upper left corner of the screen to return to the ‘Administrator Page’ screen. Importing and Deleting Methods As Promega provides new purification kits for use with the Maxwell CSC 48 instrument, new ®...
  • Page 37 Methods can be deleted from the software by touching the X on the right side of the method list item. Import new methods using the Select Import Package button. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 38 'Preferences' screen. To leave the 'Preferences' screen without saving changes, touch the Don't Save button. Touch the Cancel button to return to the 'Preferences' screen without saving changes. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 39 'Preferences' screen. To leave the 'Preferences' screen without saving changes, touch the Don't Save button. Touch the Cancel button to return to the 'Preferences' screen without saving changes. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 40 • SSL Encrypted Connection: Checkbox indicating whether the e-mail account uses an SSL Encrypted connection Note: Contact your IT department to provide the information required to complete e-mail settings. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 41 Use the Test Connection button to check the settings entered on this screen. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 42 3. Once the desired file path has been specified, touch the OK button to save the new export folder setting or touch the Cancel button to cancel any changes. After touching either OK or Cancel, you will be returned to the ‘Audit Records’ screen. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 43 Windows Operating System. After saving, an ® informational screen will be displayed indicating that “Windows will now restart”. Touch the OK button to restart the operating system. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 44 ® Windows Operating Figure 31. ‘Instrument Name’ screen. Use this screen to manually enter a name for this System. Maxwell CSC 48 Instrument. ® Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 45: Operating The Maxwell ® Csc 48 Instrument

    1. Prepare samples for nucleic acid extraction following the instructions provided in the Technical Manual for the specific Maxwell CSC reagent kit. From the ‘Home’ screen ® (Figure 32), select the Start button to begin the process of running an extraction method. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 46 Back button to return to the ‘Home’ screen. Scanning or entering the bar code automatically selects the protocol to be run. Figure 33. ‘Scan Barcode’ screen. Scanning the method bar code automatically selects the protocol on the Maxwell CSC 48 Instrument. ® Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 47 Upon pressing the OK button on the user prompt, the software will return to the ‘Home’ screen. Please contact Promega Technical Services (techserv@promega.com) to get the most up-to-date information on the available methods.
  • Page 48 Note that positions with incomplete information will display a red circle with an exclamation point at the top of the gray rectangle. The Proceed button will be gray and inactive while required information is missing. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 49 To select/deselect a cartridge position touch the long rectangle for any given position. Use the Front and Back buttons to toggle between views of the front (positions 1–24) and back (positions 25–48) deck trays. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 50 Proceed button to display the ‘Door’ prompt (Figure 37). Press the OK button to open the Maxwell CSC 48 Instrument door. ® Figure 37. ‘Door’ screen. Informs the user that the Maxwell CSC 48 Instrument door will ® open. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 51 Figure 39. Placing the cartridges in the Deck Tray and pressing firmly to snap them in place. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 52 Ensure the deck tray is level on the instrument platform and fully seated. After all of these steps have been performed, touch the Start button to begin the purification, or touch the Cancel button to return to the previous screen. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 53 ‘Running’ screen while a method is running. If you wish to abort the current run, touch the Abort button in the lower right corner of the screen. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 54 (Figure 43). After a method is complete the Current Step will be listed as Completed. Touch the Open Door button to open the door of the Maxwell ® 48 Instrument. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 55 Do not reuse reagent cartridges, plungers or elution tubes. Figure 44. Closing the elution tube caps. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 56 UV sanitization (see Section 6.6, Sanitizing). Figure 46. ‘Report View’ screen. A run report is displayed after the completion of an extraction run. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 57 After a method is aborted, the Current Step will list the reason for aborting the method. error) will result in the loss of all the samples. Do not attempt to repurify samples from an aborted run. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 58: Clean Up

    Touch the Start Clean Up button to eject the remaining plungers. After the Clean Up is successful, you can press the Open Door button and remove the deck tray. If the plunger Clean Up fails, you should contact Promega Technical Services for further assistance. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 59: Results

    (Completed or Aborted), the time the method run was started, the length of time the method took to process, the operator ID and additional instrument information. Buttons are present on the left side of the screen to Print and Export report information. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 60 You can use the tab-separated format file (*.txt) with Laboratory Information Management Systems (i.e., LIMS). An example of the tab-separated format file is shown in Figure 51 and an example of the printable PDF format is shown in Figure 52. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 61 SAM04 321654 2020-12 AS1780 SAM05 321654 2020-12 AS1780 SAM06 321654 2020-12 AS1780 SAM07 321654 2020-12 AS1780 SAM08 321654 2020-12 Figure 52. Example of a report in the PDF format. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 62: Running Reports

    Figure 53. Maxwell CSC 48 ‘Home’ screen. Selecting Results opens the Maxwell CSC 48 ® ® ‘Results’ screen from which extraction reports from all method runs can be viewed. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 63 Press the Export All button to export all displayed reports to a user-specified drive location. Figure 55. Extraction report. An example of the sample tracking information present in an extraction report. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 64: Sanitizing

    Figure 57. ‘Sanitize Checklist’ screen. The sanitization checklist indicates how long the UV sanitization will take. Confirm that no samples or eluates are present in the instrument prior to running Sanitization. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 65 UV treatment is not a substitute for cleaning. Using the UV Sanitization protocol alone may not provide sufficient decontamination. Follow the cleaning guidelines in Section 7. Figure 58. Example Sanitization Report. The sanitization report will be displayed after completion of sanitization. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 66 Operating the Maxwell® CSC 48 Instrument Notes Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 67: Cleaning And Maintenance

    • Never allow liquids to remain on instrument surfaces for extended periods of time. accordance with your institutional • Keep all moisture away from the heated elution tube slots to prevent damaging the guidelines. heating elements. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 68: Cleaning The Hardware

    Removal of paramagnetic particles from the magnetic rod assembly will require multiple wipes. The damp cloth may be wrapped around a magnet to facilitate removal of magnetic particles. 7. If the magnetic rod assemblies cannot be cleaned, please contact Promega Technical Services for assistance.
  • Page 69: Dealing With Spills

    Maxwell reagent ® cartridges, and should not be added to any sample waste containing lysis solutions. Bleach should not be used to clean reagent spills. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 70 Cleaning and Maintenance Notes Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 71: Troubleshooting

    Troubleshooting For questions not addressed here, please contact your local Promega Branch Office or Distributor. Contact information available at: www.promega.com. E-mail: techserv@promega.com Symptoms Causes and Comments Clean Up function If plungers are still engaged on the plunger bars, perform the following steps: does not unload the •...
  • Page 72 Services. Open door detected during Open door was detected during operation. The run will be aborted operation and samples will be lost. If error persists, contact Promega Technical Services. Protocol: Aborted By User User aborted the method. An aborted run (initiated by the user) will result in the loss of all the samples.
  • Page 73: Use Of A Usb Flash Drive

    User warning informing user that a firmware version change has been Firmware version change detected detected. Contact Promega Technical Services. 8.1 Use of a USB Flash Drive • When using a USB flash drive, insert it before running any method and do not remove it until the method is finished.
  • Page 74 Notes Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 75: Instrument Service

    CSC 48 Instrument is designed to provide years of consistent performance with ® little maintenance. If a problem arises with your instrument, please contact Promega or your local Promega representative for support. Visit the Promega Web site at www.promega.com for the contact information for the Promega branch or distributor nearest you.
  • Page 76: Repacking The Maxwell ® Csc 48 Instrument

    CSC 48 Instrument ® Note: If you do not have the original Maxwell CSC 48 Instrument packaging, please contact ® Promega Technical Services or your local Promega representative to order Maxwell CSC 48 ® Instrument packaging. Ship the Maxwell CSC 48 Instrument only with Promega packaging to avoid any damages.
  • Page 77 6. Secure the four shipping screws by screwing them in on both sides of the magnet and plunger bars as shown in Figure 62. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 78 9. Place the instrument in the bottom foam packaging material. The instrument will only fit properly into the bottom foam piece in one orientation (Figure 64). Figure 64. Maxwell CSC 48 Instrument in correct orientation in the box. ® Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 79 ® 12. Close the flaps on the shipping container top, and secure them with packing tape. 13. Write the return authorization number provided to you by Promega or your local Promega representative on the outside of the shipping box. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 80: Instrument Disposal

    Instrument Service 9.4 Instrument Disposal Contact your local Promega Representative for disposal of the instrument. Please follow your institutional and country-specific requirements to handle the disposal of accessories. Instrument must be decontaminated prior to disposal. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20...
  • Page 81: Warranties, Service Agreements And Related Products

    We will repair it and return it performing to original factory specifications. PM visits are available separately. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 82 This service product must be delivered by a Promega representative who is certified to perform the Installation Qualification. The service product involves a site visit to perform: •...
  • Page 83 Promega at Promega’s expense. Promega will return the repaired or replaced unit to you at Promega’s expense within 3 business days after the repair is complete. This Warranty may be renewed, in one-year terms, provided the renewal is requested prior to the expiration of the current Warranty or Service Agreement.
  • Page 84: Related Products

    Maxwell RSC/CSC 48 Back Deck Tray 1 each AS8402 ® USB Ethernet Adapter 1 each AS8403 For In Vitro Diagnostic Use. This product is only available in certain countries. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 85: Certificate Of Decontamination

    ** The signature of a Radiation Safety Officer is also required if the instrument was used with radioactive materials. This instrument is certified by the undersigned to be free of radioactive contamination. Date: _____________________________________________________________________________ Place: ____________________________________________________________________________ Signature: _________________________________________________________________________ Name (block capital letters): __________________________________________________________ Title: _____________________________________________________________________________ Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 86 Notes Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 87: Summary Of Change

    Summary of Change The following change was made to the 6/20 revision of this document: Updated Methods link in Section 5.3. Promega Maxwell CSC 48 Instrument | TM623 | Revised 6/20 ®...
  • Page 88 Products may be covered by pending or issued patents or may have certain limitations. Please visit our Web site for more information. All prices and specifications are subject to change without prior notice. Product claims are subject to change. Please contact Promega Technical Services or access the Promega online catalog for the most up-to-date information on Promega products.
  • Page 89 PROMEGA CORPORATION 2800 WOODS HOLLOW ROAD MADISON, 53711-5399 TELEPHONE 608-274-4330 • • • www.promega.com ©2020 PROMEGA CORPORATION ALL RIGHTS RESERVED PRICES AND SPECIFICATIONS SUBJECT TO CHANGE WITHOUT PRIOR NOTICE TM623 REVISED 6/20 • • • • •...