Reverse Purge (Retrieving Unused Reagents from Injectors, optional)
1.
Tap the Reverse Purge button on the Injectors screen (Figure 94). The door will open.
Figure 96. Injector Purge function.
2.
Tap the Start Reverse Purge button (Figure 96).
3.
Any reagents left in the fluid lines will be returned to the reagent container.
Flushing the Injectors after Use
1.
Remove the sample plate from the instrument.
2.
Tap the Flush button on the 'Injectors' screen (Figure 94). The door will open.
Figure 97. Injector Flush function.
3.
Place the Waste Collection Tray on the Microplate Sample Tray.
4.
Tap the Start Flush button (Figure 97).
The recommended flush protocol uses a sequence of 1 cycle (1ml) of deionized water, 1 cycle (1ml) of 70%
ethanol, 1ml of deionized water and then 1 cycle (1ml) of air to flush the fluid.
Note: Remove Waste Collection Tray when priming and flushing are complete. See Section 5.D for injector
cleaning instructions.