Symptoms
Low luminescent signal
High luminescent cross talk
6.C. Use of a USB Flash Drive
Insert a USB flash drive before running any protocol and do not remove until the protocol is finished.
Do not insert or remove USB flash drive while instrument is running.
Because USB flash drives may vary from vendor to vendor or from type to type, incompatibilities may occur.
If the USB flash drive is not detected after a few seconds or if you encounter any issues with the USB flash
drive, shut down and restart the Tablet PC and the instrument, and try a different brand of USB flash drive.
If the Tablet PC becomes unresponsive after inserting a USB flash drive, shut down and restart the Tablet
PC and the instrument.
7.

Appendix

7.A. Configuring the Tablet PC
The GloMax
Discover Tablet PC touch screen allows the user to make selections from a series of options.
®
Selections may be made by tapping the appropriate option on the Tablet PC screen.
Your IT department or site Administrator should configure the Tablet PC according to the IT rules and IT procedures
pertinent to your institution. The following is a set of guidelines for your IT personnel or site Administrator.
Depending on your institution's IT policy, you may or may not be able to configure/change the settings for the Tablet
PC. The Tablet is pre-installed with a Microsoft Excel
if you desire. We do not recommend loading other programs onto the Tablet because these may interfere with the
instrument operation.
Note: Some institutions require anti-virus software installed on PCs within the institution. If installing an
!
antivirus program please ensure it is set to manual, not automatic, update.
Antivirus software will slow the performance of the Tablet PC and will affect the performance of the GloMax
Instrument while it is measuring samples.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
www.promega.com
Causes and Comments
Low activity in sample.
Use white opaque plates.
Use high-quality white or black opaque plate.
If using 384-well plates, confirm that the 384-well aperture
has been installed.