3. Select and drag the "User Prompt" icon to the Protocol Composer and
type an instructional message into the Edit bar (Figure 12).
4. Select and drag the "Fluorescence" icon to the Protocol Composer and
select the Red Optical Kit (Figure 13).
5. Insert a plate into the GloMax
"Start" button to begin a run.
Figure 13. A Defined Protocol ready to run.
7.
GloMax
®
-Multi+ Detection System with Shaking
The GloMax
®
-Multi+ Detection System with Shaking allows the sample plate to
be shaken as is necessary to meet the requirements of certain assays. Selecting
the "Shaking" icon allows one to define the parameters of shake operation such
as time, intensity and type of motion for the shaking pattern.
Time: The GloMax
®
-Multi+ Detection System with Shaking allows for shaking
intervals of 0.1–120 minutes.
Intensity: The GloMax
of defined shaking intensity: low, medium and high. These settings roughly
equate to 150, 300 and 500 rpm, respectively.
Type of motion: Either linear, where the entire platform of the shaker moves in
a linear motion or orbital, where the entire platform of the shaker moves in a
circular orbit.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Printed in USA.
Revised 5/10
®
-Multi+ Detection System and press the
®
-Multi+ Detection System with Shaking has three rates
Part# TM311
Page 19