Acquisition Specifications; Data Specifications; Table 74 Acquisition Specifications - Omni-M2-Ut-2C; Table 75 Data Specifications - Omni-M2-Ut-2C - Olympus OmniScan MX User Manual

Multitechnology flaw detector
Hide thumbs Also See for OmniScan MX:
Table of Contents

Advertisement

DMTA-20015-01EN [U8778402], Rev. M, February 2017
O.2

Acquisition Specifications

This section lists the acquisition specifications for frequency, data display, and
synchronization (see Table 74 on page 198).
.
Table 74 Acquisition specifications — OMNI-M2-UT-2C
Digitizing frequency
A-scan (acquisition
mode)
Maximum pulsing rate
Depth in material
Refresh rate
Envelope (echo-dynamic
mode)
On internal clock
External pace
On encoder
O.3

Data Specifications

This section details the data specifications of processing, TCG, and storage (see
Table 75 on page 198).
Number of data points
Real-time averaging
198
Appendix O
100 MHz
Up to 6000 A-scans per second (512-point,
8-bit A-scan)
Up to 10 kHz (C-scan)
59.8 meters in steel (longitudinal wave),
10 ms with compression
0.49 meter in steel (longitudinal wave),
81.9 µs without compression
A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz
Yes: Volume-corrected S-scan (30 Hz)
Synchronization
1 Hz to 10 kHz
Yes
On 2 axes: from 1 to 65536 steps
Table 75 Data specifications — OMNI-M2-UT-2C
Up to 8192
2, 4, 8, 16, 32, 64
Frequency
Display